EasyManua.ls Logo

ConMed Advantage - 3.2 CoolFlex High Speed Drill (E9010) Operation

ConMed Advantage
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONTROLLER
OPERATION
56
A
(b) When operating in ON/OFF mode, the
handpiece motor is either ON (pedal
depressed) or OFF (pedal released). The
motor rotates at a fixed rate regardless
of the amount of pressure applied to the
pedal (Reference 2.1.5 Controller
Operation” on page 48, “3.6.5 Group
2 Handpiece Cord Button
Operation” on page 93 or “3.9.2.4
Using the Handpiece Cord Button to
Change Speed” on page 100 for
information to change operating
speeds).
2.1.7.1 Operation of the 2-Pedal
Footswitch (5020-053)
Operation of the footswitch when using a
MicroChoice Sagittal Saw (5020-022),
Reciprocating Saw (5020-023), Oscillator
Handpiece (5020-024), or E9015 Cranial
Perforator Drive:
1. Depress either the left or right pedal to
activate the handpiece.
NOTE: There is no oscillate or reverse
mode for these handpieces.
Operation of the footswitch when using a
MicroChoice Elite High Speed Drill
(5020-025), Medium Speed Drill (5020-021),
or Low Speed Drill (5020-026) and the
CoolFlex High Speed Drill (E9010):
1. Depress the right pedal to activate
the handpiece in the forward
direction.
2. Depress the left pedal to activate
the handpiece in the reverse
direction. A brief audible beeping
tone will indicate reverse direction when
the handpiece is initially activated.

Table of Contents