79
MICROCHOICE
HANDPIECES
C
3.4.3.5 Laminectomy Bur Guard
(5020-067)
NOTES:
1. Laminectomy burs are the only burs
inserted prior to placing a guard or an
attachment on the handpiece.
2. For additional information refer
to the MicroChoice Angled
Attachment Package Insert or
the MicroChoice Bur Guard Package
Insert.
1. To fasten the bur to the handpiece:
(a) Rotate the handpiece bur lock collar
counterclockwise to the “OPEN
position.
(b) Insert the correct bur (5092-103
shown).
(c) Rotate the handpiece bur lock collar
clockwise until the indicator dots are
aligned in the “LOCK” position to lock
the bur into place.
(d) Pull on the bur to ensure it is locked in
place.
2. To fasten the guard to the handpiece:
(a) Slide the Laminectomy Guard over the
bur and the end of the drill. Be certain
the guard is completely seated.
3. To remove the guard:
(a) Pull on the guard to remove it from the
handpiece.
4. To remove the bur:
(a) Rotate the handpiece bur lock collar
counterclockwise to the “OPEN
position and remove the bur.