Curtiss-Wright | PG Drives Technology
R-net Omni2 Technical Manual | Chapter 8 Diagnostics | 131
SK81935-01
3.1 CENTER JOYSTICK
The most common cause of this trip is if the SID is deflected away from center before and during the time the control system
is switched on. The joystick-displaced screen will be displayed for 5 seconds. If the SID is not released within that time then a
trip is registered. Although a trip screen is not displayed, the system log will show the trip and numbers of occurrences.
Ensure that the SID is centered and power-up the control system.
If the trip is still present then the SID or Omni2 may be defective. Refer to Section 5.
3.2 JOYSTICK CALIBRATION ERROR
This occurs when the Joystick-Type SID Calibration process has not been successful.
Enter OBP and attempt calibration.
If the trip is still present after the above has been attempted, then the Omni2 may be defective. Refer to Section 5.
3.3 LATCHED TIMEOUT
This occurs when the control system detects that the programmed Latched Timeout period has been exceeded. For example,
the SID Joystick, Head Array, Sip and Puff etc. has not been operated frequently enough.
The trip is a notification of why the control system has dropped-out of Latched Mode.
Cycle the power.
Initiate Latched Mode.
If the trip is still present after the above checks have been made, then the SID may be defective. Refer to Section 5.
3.4 DIME ERROR
This occurs when the control system detects an identification conflict between two modules in the system.
If a new module has been introduced:
Disconnect the new module and cycle the power.
If no trip is present connect the new module to the system and cycle the power.
If the trip reappears then the new module must be the cause of the problem.
If there have been no additions:
Disconnect one module at a time and cycle the power.
If the trip is still present after the above checks have been made, contact your service agent.