Cato
edition
Anaesthetic Workstation
Instructions for Use
Software versions:
Ventilator: ...................... 7.n
Monitor: ......................... 2.n
MT-439-2000
D

Table of Contents

Questions and Answers

  • C
    Cindy HopkinsSep 8, 2025
    What causes the FRESH GAS ? alarm on Dräger Cato, and how can I resolve it?
    • R
      rogerraySep 8, 2025
      If fresh gas in manual ventilation bag is too little, increase the fresh gas flow and repair any leaks in the hose system.
  • J
    Jordan SpencerSep 10, 2025
    What does PAW NEGATIVE mean on my Dräger Medical Equipment and how do I fix it?
    • C
      casey72Sep 10, 2025
      If a negative airway pressure is measured, check the hose system and tube on the ventilator.
  • W
    wgarzaSep 12, 2025
    How do I fix the % O2 ERR alarm on my Dräger Cato?
    • J
      jeffery36Sep 12, 2025
      If the sensor for inspiratory O2 measurement is incorrectly calibrated, calibrate the sensor. If the sensor was changed and/or not calibrated, replace and calibrate the cell. If the sensor is not plugged in, plug in the sensor connector. If the sensor lead is defective, replace the sensor housing. If a non-Dräger sensor cell is used, use an original sensor cell.
  • B
    Brandon FlowersSep 13, 2025
    What does AGT NOT SEL mean on Dräger Cato Medical Equipment and how to resolve it?
    • S
      smithkarenSep 13, 2025
      If the anaesthetic agent is not selected or the selection differs from that detected by Vapor identification, check and select the anaesthetic agent.
  • J
    jacob36Sep 15, 2025
    How to fix INSP CO2 HI alarm on Dräger Cato?
    • K
      Kara WellsSep 15, 2025
      If the inspiratory CO2 concentration is above the upper alarm limit of 5 mmHg, replace the soda lime in the circle system of the anaesthetic workstation.
  • R
    Richard WellsSep 18, 2025
    What to do if CO2 LINE BLK alarm appears on Dräger Cato?
    • J
      James BrownSep 18, 2025
      If the sample line is blocked, replace the sample line and water trap if necessary.

Dräger Cato Specifications

General IconGeneral
TypeAnesthesia Workstation
Tidal Volume Range20 to 1400 ml
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
Integrated MonitoringYes
Power Supply100 - 240 V AC, 50/60 Hz
UsageOperating Room
Ventilation ModesVolume Controlled
Vaporizer MountsUp to 2
AlarmsGas Supply Failure