Object name Command Parameter range Factory set-
ting
Data type Units Comments
Normal speed seng !S856 50..100 80 decimal % Normal speed as a percentage of full
speed
?S856 - - - - -
Standby speed set-
ting
!S857 55..100 70 decimal % Standby speed as a percentage of full
speed
?S857 - - - - -
Temperature read-
ings
?V859 0..100 - decimal °C Measured motor temperature
0..100 decimal °C Measured motor controller temperature
Link parameter read-
ings
?V860 0..500 - decimal 0.1 Volts Measured link voltage
0..300 decimal 0.1 Amps Measured link current
0..15000 decimal 0.1 Was Measur
ed link power
Pump run hours ?V862 0..65535 - decimal hours Hours run by pump
Temperature read-
ings 2
?V865 -200..200 - decimal °C Measured motor temperature
-200..200 decimal °C Measured motor controller temperature
-200..200 decimal °C Measured bearing temperature
Factory sengs !S867 1 - - - Reset all conguraon opons and pa-
rameters to the factory sengs
PIC soware version ?S868 10 - string chars Boot loader soware version number
(D396696XXX)
Speed control !C869 0 - decimal - Set target speed to full speed
1 - - - Set target speed to standby speed
Timer opons !S870 0 1 decimal - Timer = disabled
?S870 1 - - - Timer = enabled
Note that the mer is permanently ena-
bled on ramp-up.
08/2021 - ©Edwards Limited
Page 40B8J200880_D
300812671_002_C3
B8J200880_D - Technical data