Object name Command Parameter range Factory set-
ting
Data type Units Comments
Analogue signal op-
ons
!S871 0 0 decimal - Analogue output = measured speed
?S871 1 - Analogue output = measured power
2 Analogue output = measured motor
temp.
3 Analogue output = measur
ed control
temp.
4 Analogue output = measured bearing
temp.
Electronic braking
opons
!S872 0 0 decimal Electronic braking = disabled
?S872 1 Electronic braking = enabled. Refer to
Electronic braking on page 17
Close vent valve !C875 1 - decimal - Closes the vent valve for delayed start
and overrides the current vent opon.
There is no open vent valve command
but the stop command (!C852 0) will
clear the override.
Valve 1 type !S877 0 0 decimal - Normally Open Valve
?S877 1 Normally Closed Valve
(power cycle is required)
Valve 2 type !S878 0 0 decimal - Normally Open Valve
?S878 1 Normally Closed Valve
(power cycle is required)
Service status ?V881 32 bits - hex ags Service status word
Controller run me ?V882 0..999999 - decimal hours Hours run by controller
0..999999 decimal hours Hours unl con
troller service due
08/2021 - ©Edwards Limited
Page 41B8J200880_D
300812671_002_C3
B8J200880_D - Technical data