Object name Command Parameter range Factory set-
ting
Data type Units Comments
Pump run me ?V883 0..999999 - decimal hours Hours run by pump
0..999999 decimal hours Hours unl pump service due
Pump cy
cles ?V884 0..65535 - decimal cycles Cycles run by pump
0..65535 decimal cycles Cycles unl pump service due
Bearing run me ?V885 0..999999 - decimal hours Hours run by bearing
0..999999 decimal hours Hours unl bearing service due
Vent opons 1 !S853
0
0 decimal
Hard vent when < 50% speed if stop or
fail
1 Controlled vent if > 50% speed then
hard vent if
< 50% speed if stop or fail
2 If stop, hard vent
If fail, hard vent when < 50% speed
3 If stop, hard vent
If fail, controlled vent > 50% speed then
hard vent < 50% speed
4 If fail, hard vent
If stop, hard vent when
< 50% speed
5 If fail, hard vent
If stop, controlled vent > 50% speed
then hard vent < 50% speed
6 Hard vent if stop or fail
7 Same as opon 6
8 Output is permanently energised (Fan
ON)
08/2021 - ©Edwards Limited
Page 42B8J200880_D
300812671_002_C3
B8J200880_D - Technical data