Object name Command Parameter range Factory set-
ting
Data type Units Comments
9 Do not use
10 Output is permanently de-energised
(Fan OFF)
11 Do not use
12 Do not use
13 Do not use
14 Har
d vent when < 50% speed if stop
15 Controlled vent if > 50% speed then
hard vent if < 50% speed if stop
Vent opons 2 ?864
0
8 decimal
Hard vent when < 50% speed if stop or
fail
1 Controlled vent if > 50% speed then
hard vent if < 50% speed if stop or fail
2 If stop, hard vent
If fail, hard vent when < 50% speed
3 If stop, hard vent
If fail, controlled vent > 50% speed then
hard vent < 50% speed
4 If fail, hard vent
If stop, hard vent when < 50% speed
5 If fail, hard vent
If stop, controlled vent > 50% speed
then hard vent < 50% speed
6 Hard vent if stop or fail
7 Same as opon 6
8 Output is permanently energised (Fan
ON)
08/2021 - ©Edwards Limited
Page 43B8J200880_D
300812671_002_C3
B8J200880_D - Technical data