Operating menu and parameter description iTEMP TMT162
116 Endress+Hauser
User entry • Alarm
• Warning
• Disabled
Factory setting For detailed information see the 'Overview of diagnostic events' →
37
Status signal
Navigation Expert → Diagnostics → Diagnostic settings → Status signal
Description A certain status signal is assigned at the factory to each diagnostic event in the categories:
sensor, electronics, process and configuration
1)
. The user can change this assignment
for certain diagnostic events via the diagnostic settings. →
37
1) Digital information available via HART® communication
User entry • Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• No effect (N)
Factory setting For detailed information see the 'Overview of diagnostic events' → 37