EasyManua.ls Logo

Endress+Hauser iTEMP TMT162 - Setup Menu

Endress+Hauser iTEMP TMT162
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iTEMP TMT162 Operating menu and parameter description
Endress+Hauser 69
14.1 "Setup" menu
This menu contains all the parameters that are needed to configure the basic settings of
the device. The transmitter can be put into operation with this limited parameter set.
n = Stands for the number of sensor inputs (1 and 2)
Device tag
Navigation Setup → Device tag
Diagnostics → Device information → Device tag
Expert → Diagnostics → Device information → Device tag
Description Use this function to enter a unique name for the measuring point so it can be identified
quickly within the plant. This name is shown on the display.
User entry Max. 32 characters, such as letters, numbers or special characters (e.g. @, %, /)
Factory setting 32 x '?'
Unit
Navigation Setup → Unit
Expert → System → Unit
Description Use this function to select the engineering unit for all the measured values.
Options °C
°F
K
°R
Ohms
mV
Factory setting °C
Sensor type n
Navigation Setup → Sensor type n
Expert → Sensor → Sensor n → Sensor type n
Description Use this function to select the sensor type for the sensor input in question.
Sensor type 1: settings for sensor input 1
Sensor type 2: settings for sensor input 2
Please observe the terminal assignment when connecting the individual sensors. In
the case of 2-channel operation, the possible connection options must also be
observed.

Table of Contents

Other manuals for Endress+Hauser iTEMP TMT162

Related product manuals