Commissioning Proline Prosonic Flow P 500 HART
134 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Network security Select the security type of the
WLAN network.
Unsecured
WPA2-PSK
EAP-PEAP with
MSCHAPv2
*
EAP-PEAP
MSCHAPv2 no
server authentic.
*
EAP-TLS
*
WPA2-PSK
Security identification Select security settings and
download these settings via
menu Data management >
Security > WLAN.
Trusted issuer
certificate
Device certificate
Device private key
User name Enter user name.
WLAN password Enter WLAN password.
WLAN IP address Enter IP address of the WLAN
interface of the device.
4 octet: 0 to 255 (in
the particular octet)
192.168.1.212
WLAN MAC address Enter MAC address of the
WLAN interface of the device.
Unique 12-digit
character string
comprising letters
and numbers
Each measuring
device is given an
individual address.
WLAN passphrase The WPA2-PSK option is
selected in the Security type
parameter.
Enter the network key (8 to 32
characters).
The network key
supplied with the device
should be changed
during commissioning
for security reasons.
8 to 32-digit
character string
comprising numbers,
letters and special
characters (without
spaces)
Serial number of the
measuring device
(e.g. L100A802000)
Assign SSID name Select which name will be used
for SSID: device tag or user-
defined name.
Device tag
User-defined
User-defined
SSID name The User-defined option is
selected in the Assign SSID
name parameter.
The WLAN access point
option is selected in the
WLAN mode parameter.
Enter the user-defined SSID
name (max. 32 characters).
The user-defined SSID
name may only be
assigned once. If the SSID
name is assigned more
than once, the devices
can interfere with one
another.
Max. 32-digit
character string
comprising numbers,
letters and special
characters
EH_device
designation_last 7
digits of the serial
number (e.g.
EH_Prosonic_Flow_5
00_A802000)
Connection state Displays the connection status. Connected
Not connected
Not connected
Received signal strength Shows the received signal
strength.
Low
Medium
High
High
* Visibility depends on order options or device settings
10.5.6 Configuration management
After commissioning, you can save the current device configurationor restore the previous
device configuration. The device configuration is managed via the Configuration
management parameter.

Table of Contents