Commissioning Proline Prosonic Flow P 500 HART
120 Endress+Hauser
Failure mode
→ 120
Invert output signal
→ 120
Parameter overview with brief description
Parameter Description Selection / User interface /
User entry
Factory setting
Signal mode Select the signal mode for the double pulse
output.
• Passive
• Active
*
• Passive NE
Passive
Master terminal number Shows the terminal numbers used by the
master of the double pulse output module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Assign pulse output Select process variable for pulse output. • Off
• Volume flow
• Mass flow
• GSV flow
*
• NSV flow
*
• S&W volume flow
*
• Product 1 volume flow
*
• Product 2 volume flow
*
• Product 3 volume flow
*
• Product 4 volume flow
*
• Product 5 volume flow
*
• Product 1 mass flow
*
• Product 2 mass flow
*
• Product 3 mass flow
*
• Product 4 mass flow
*
• Product 5 mass flow
*
• Product 1 NSV flow
*
• Product 2 NSV flow
*
• Product 3 NSV flow
*
• Product 4 NSV flow
*
• Product 5 NSV flow
*
Off
Measuring mode Select measuring mode for pulse output. • Forward flow
• Forward/Reverse flow
• Reverse flow
• Reverse flow compensation
Forward flow
Value per pulse Enter measured value at which a pulse is
output.
Signed floating-point number Depends on country and
nominal diameter
Pulse width Define time width of the output pulse. 0.5 to 2 000 ms 0.5 ms
Failure mode Select output behavior in the event of a
device alarm.
• Actual value
• No pulses
No pulses
Invert output signal Invert the output signal. • No
• Yes
No
* Visibility depends on order options or device settings
10.4.12 Configuring the local display
The Display wizard guides you systematically through all the parameters that can
configured for configuring the local display.