Parameter overview with brief description
Parameter Description User interface
Signal strength Displays the current signal strength (0 to 100 dB).
Assessment of the signal strength:
• < 10 dB: bad
• > 90 dB: very good
Signed floating-point number
Acceptance rate Displays the ratio of the number of ultrasonic signals accepted
for flow calculation and the total number of ultrasonic signals
emitted.
0 to 100 %
Signal to noise ratio Displays the current signal to noise ratio (0 to 100 dB).
Assessment of the signal-to-noise ratio:
• < 20 dB: bad
• > 50 dB: very good
Signed floating-point number
Turbulence Displays the current turbulence. Signed floating-point number
11.4.3 "Input values" submenu
The Input values submenu guides you systematically to the individual input values.
Navigation
"Diagnostics" menu → Measured values → Input values
‣
Input values
‣
Current input 1 to n
→ 150
‣
Status input 1 to n
→ 151
Input values of current input
The Current input 1 to n submenu contains all the parameters needed to display the
current measured values for every current input.
Navigation
"Diagnostics" menu → Measured values → Input values → Current input 1 to n
‣
Current input 1 to n
Measured values 1 to n
→ 150
Measured current 1 to n
→ 150
Parameter overview with brief description
Parameter Description User interface
Measured values 1 to n Displays the current input value.
Dependency
The unit is taken from the Pressure unit parameter
Signed floating-point number
Measured current 1 to n Displays the current value of the current input. 0 to 22.5 mA