Loading...
Proline Prosonic Flow P 500 HART Diagnosis and troubleshooting
Endress+Hauser 177
12.8 Pending diagnostic events
The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
Via local display →  165
Via web browser →  167
Via "FieldCare" operating tool →  169
Via "DeviceCare" operating tool →  169
Other pending diagnostic events can be displayed in the Diagnostic list submenu
→  177
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
→  177
Previous diagnostics
→  177
Operating time from restart
→  177
Operating time
→  177
Parameter overview with brief description
Parameter Prerequisite Description User interface
Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic
event along with its diagnostic
information.
If two or more messages occur
simultaneously, the message with
the highest priority is shown on
the display.
Symbol for diagnostic
behavior, diagnostic code
and short message.
Previous diagnostics Two diagnostic events have already
occurred.
Shows the diagnostic event that
occurred prior to the current diagnostic
event along with its diagnostic
information.
Symbol for diagnostic
behavior, diagnostic code
and short message.
Operating time from restart Shows the time the device has been in
operation since the last device restart.
Days (d), hours (h),
minutes (m) and seconds
(s)
Operating time Indicates how long the device has been
in operation.
Days (d), hours (h),
minutes (m) and seconds
(s)
12.9 Diagnostics list
Up to 5 currently pending diagnostic events can be displayed in the Diagnostic list
submenu along with the associated diagnostic information. If more than 5 diagnostic
events are pending, the events with the highest priority are shown on the display.
Navigation path
Diagnostics → Diagnostic list