7.6 Ensuring the degree of protection
The measuring device fulfills all the requirements for the degree of protection IP66/67,
Type 4X enclosure.
To guarantee the degree of protection IP66/67, Type 4X enclosure, carry out the following
steps after electrical connection:
1. Check that the housing seals are clean and fitted correctly.
2. Dry, clean or replace the seals if necessary.
3. Tighten all housing screws and screw covers.
4. Firmly tighten the cable glands.
5. To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
A0029278
6. The cable glands supplied do not ensure housing protection when not in use. They
must therefore be replaced by dummy plus corresponding to the housing protection.
7.7 Post-connection check
Are cables or the device undamaged (visual inspection)?
Is the protective earthing established correctly?
Do the cables used comply with the requirements ?
Are the mounted cables relieved of tension?
Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap"
→ 59?