Proline Prosonic Flow P 500 HART Technical data
Endress+Hauser 201
Galvanic isolation The outputs are galvanically isolated:
• from the power supply
• from one another
• from the potential equalization (PE) terminal
DN 50 to 4000 (2 to 160") and non-hazardous area: The clamp-on sensors can also be
mounted on cathodically protected pipes. Solution available on request. Not applicable to
order code for "Sensor version", options AG, AH.
Protocol-specific data
Manufacturer ID 0x11
Device type ID 0x5D (93)
HART protocol revision 7
Device description files
(DTM, DD)
Information and files under:
www.endress.com
HART load Min. 250 Ω
System integration Information on system integration → 89.
• Measured variables via HART protocol
• Burst Mode functionality
16.5 Power supply
Terminal assignment → 47
Supply voltage
Order code
"Power supply"
Terminal voltage Frequency range
Option D DC 24 V ±20% –
Option E AC 100 to 240 V –15…+10% 50/60 Hz, ±4 Hz
Option I
DC 24 V ±20% –
AC 100 to 240 V –15…+10%
• 50/60 Hz
• 50/60 Hz, ±4 Hz
Power consumption Transmitter
Max. 10 W (active power)
switch-on current Max. 36 A (<5 ms) as per NAMUR Recommendation NE 21
Current consumption Transmitter
• Max. 400 mA (24 V)
• Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz)
Power supply failure • Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memory or
in the pluggable data memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.