EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Parameter Overview with Brief Description; Operating Mode

Endress+Hauser Proline Prosonic Flow P 500
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow P 500 HART Commissioning
Endress+Hauser 113
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Operating mode Define the output as a pulse,
frequency or switch output.
Pulse
Frequency
Switch
Pulse
Terminal number Shows the terminal numbers
used by the PFS output
module.
Not used
24-25 (I/O 2)
22-23 (I/O 3)
Signal mode Select the signal mode for the
PFS output.
Passive
Active
*
Passive NE
Passive
Assign frequency output The Frequency option is
selected in Operating mode
parameter (→  110).
Select process variable for
frequency output.
Off
Volume flow
Mass flow
Flow velocity
Sound velocity
Temperature
*
Pressure
*
Density
*
Reference density
*
S&W volume flow
*
GSV flow
*
NSV flow
*
API gravity
*
API slope
*
Signal strength
*
Signal to noise
ratio
*
Acceptance rate
*
Turbulence
*
Electronics
temperature
Profile factor
*
Cross flow factor
*
Off
Minimum frequency value The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz
Maximum frequency value The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz
Measuring value at minimum
frequency
The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter measured value for
minimum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Measuring value at maximum
frequency
The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter measured value for
maximum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow P 500

Related product manuals