Proline Prosonic Flow P 500 HART Commissioning
Endress+Hauser 113
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Operating mode – Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
Pulse
Terminal number – Shows the terminal numbers
used by the PFS output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Signal mode – Select the signal mode for the
PFS output.
• Passive
• Active
*
• Passive NE
Passive
Assign frequency output The Frequency option is
selected in Operating mode
parameter (→  110).
Select process variable for
frequency output.
• Off
• Volume flow
• Mass flow
• Flow velocity
• Sound velocity
• Temperature
*
• Pressure
*
• Density
*
• Reference density
*
• S&W volume flow
*
• GSV flow
*
• NSV flow
*
• API gravity
*
• API slope
*
• Signal strength
*
• Signal to noise
ratio
*
• Acceptance rate
*
• Turbulence
*
• Electronics
temperature
• Profile factor
*
• Cross flow factor
*
Off
Minimum frequency value The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz
Maximum frequency value The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz
Measuring value at minimum
frequency
The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter measured value for
minimum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Measuring value at maximum
frequency
The Frequency option is
selected in the Operating
mode parameter (→  110)
and a process variable is
selected in the Assign
frequency output parameter
(→  113).
Enter measured value for
maximum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter