Technical data Proline Prosonic Flow P 500 HART
206 Endress+Hauser
Shock half-sine, according to IEC 60068-2-27
6 ms 50 g
Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21) and 43 (NE43)
Details are provided in the Declaration of Conformity.
This unit is not intended for use in residential environments and cannot guarantee
adequate protection of the radio reception in such environments.
For detailed information on high-temperature sensors CH-050 / CH-100 (order code
for "Sensor version", options AG, AH), see Special Documentation "High-temperature"
→ 217.
16.9 Process
Medium temperature range
Sensor version Frequency Temperature
C-030-A 0.3 MHz –40 to +100 °C (–40 to +212 °F)
C-050-A 0.5 MHz –20 to +80 °C (–4 to +176 °F)
C-100-A 1 MHz –20 to +80 °C (–4 to +176 °F)
C-200-A 2 MHz –20 to +80 °C (–4 to +176 °F)
C-500-A 5 MHz –40 to +150 °C (–40 to +302 °F)
C-100-B 1 MHz –40 to +80 °C (–40 to +176 °F)
C-200-B 2 MHz –40 to +80 °C (–40 to +176 °F)
C-100-C 1 MHz 0 to +170 °C (+32 to +338 °F)
C-200-C 2 MHz 0 to +170 °C (+32 to +338 °F)
CH-050-A 0.5 MHz • +150 to +220 °C (302 to +428 °F): Order code for "Process temperature", option H
• +210 to +370 °C (410 to +698 °F): Order code for "Process temperature", option I
• +350 to +550 °C (+662 to +1 022 °F): Order code for "Process temperature", option J
CH-100-A 1 MHz • +150 to +220 °C (302 to +428 °F): Order code for "Process temperature", option H
• +210 to +370 °C (410 to +698 °F): Order code for "Process temperature", option I
• +350 to +550 °C (+662 to +1 022 °F): Order code for "Process temperature", option J
Sound velocity range 600 to 3 000 m/s (1 969 to 9 843 ft/s)
Medium pressure range No pressure limitation For correct measurement, the static pressure of the medium must
be higher than the vapor pressure.
Pressure loss There is no pressure loss.
16.10 Mechanical construction
Design, dimensions For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section .