Diagnosis and troubleshooting Proline Prosonic Flow P 500 HART
182 Endress+Hauser
12.11 Resetting the measuring device
The entire device configuration or some of the configuration can be reset to a defined state
with the Device reset parameter (→  137).
12.11.1 Function range of "Device reset" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to the customer-specific value. All other parameters are reset to the factory setting.
Restart device The restart resets every parameter with data stored in volatile memory (RAM) to
the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup Restores the data that is saved on the S-DAT. Additional information: This function
can be used to resolve the memory issue "083 Memory content inconsistent" or to
restore the S-DAT data when a new S-DAT has been installed.
This option is displayed only in an alarm condition.