Proline Prosonic Flow P 500 HART Electrical connection
Endress+Hauser 49
7.3 Connecting the measuring device: Proline 500
NOTICE
An incorrect connection compromises electrical safety!
Have electrical connection work carried out by appropriately trained specialists only.
Observe applicable federal/national installation codes and regulations.
Comply with local workplace safety regulations.
Always connect the protective ground cable  before connecting additional cables.
When using in potentially explosive atmospheres, observe the information in the
device-specific Ex documentation.
7.3.1 Connecting the connecting cable
L
WARNING
Risk of damaging electronic components!
Connect the sensor and transmitter to the same potential equalization.
Only connect the sensor to a transmitter with the same serial number.
L
CAUTION
Measurement error due to shortening of the connecting cable
The connecting cable is ready for installation and must be used in the length supplied.
Shortening the connecting cable can affect the sensor's measurement accuracy.
Connecting cable terminal assignment
UP
CH1
CH2
UP
DOWN
DOWN
1.
2.
3.
4.
A0043219
1 Securing clamp
2 Connection compartment cover: sensor cable connection
3 Channel 1 upstream/downstream
4 Channel 2 upstream/downstream