EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Protecting Settings from Unauthorized Access; Write Protection Via Access Code

Endress+Hauser Proline Prosonic Flow P 500
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Proline Prosonic Flow P 500 HART
140 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry
Factory setting
Pulse value In the Pulse output
simulation parameter, the
Down-counting value option
is selected.
Set and switch off the pulse
output simulation.
0 to 65 535 0
Device alarm simulation Switch the device alarm on and
off.
Off
On
Off
Diagnostic event category Select a diagnostic event
category.
Sensor
Electronics
Configuration
Process
Process
Diagnostic event simulation Select a diagnostic event to
simulate this event.
Off
Diagnostic event
picklist (depends
on the category
selected)
Off
Current input 1 to n simulation Switch simulation of the
current input on and off.
Off
On
Off
Value current input 1 to n In the Current input 1 to n
simulation parameter, the On
option is selected.
Enter the current value for
simulation.
0 to 22.5 mA 0 mA
Status input 1 to n simulation Switch simulation of the status
input on and off.
Off
On
Off
Input signal level 1 to n In the Status input simulation
parameter, the On option is
selected.
Select the signal level for the
simulation of the status input.
High
Low
High
* Visibility depends on order options or device settings
10.7 Protecting settings from unauthorized access
The following write protection options exist in order to protect the configuration of the
measuring device from unintentional modification:
Protect access to parameters via access code →  140
Protect access to local operation via key locking →  75
Protect access to measuring device via write protection switch →  142
10.7.1 Write protection via access code
The effects of the user-specific access code are as follows:
Via local operation, the parameters for the measuring device configuration are write-
protected and their values can no longer be changed.
Device access is protected via the Web browser, as are the parameters for the measuring
device configuration.
Device access is protected via FieldCare or DeviceCare (via CDI-RJ45 service interface), as
are the parameters for the measuring device configuration.
Defining the access code via the local display
1. Navigate to the Define access code parameter (→  136).
2. Maximum of 16-digit character string comprising numbers, letters and special
characters as the access code.

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow P 500

Related product manuals