Mounting Proline Prosonic Flow P 500 HART
28 Endress+Hauser
Medium pressure range
No pressure limitation For correct measurement, the static pressure of the medium must
be higher than the vapor pressure.
6.1.4 Special mounting instructions
Weather protection cover
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0) 255 (10.0)
134 (5.3)
30 (1.18)
A0029553
21 Weather protection cover for Proline 500; engineering unit mm (in)
Cover locking: Proline 500
NOTICE
Order code "Transmitter housing", option L "Cast, stainless": The covers of the
transmitter housing are provided with a borehole to lock the cover.
The cover can be locked using screws and a chain or cable provided by the customer on
site.
‣
The use of stainless steel chains or cables is recommended.
‣
If a protective coating is applied, it is recommended to use a heat shrink tube to protect
the housing paint.
1
1
2
2
1
1
5 (0.2) min. 15 (0.6)
SW 2.5
M3
A0029799
1 Cover borehole for the securing screw
2 Securing screw to lock the cover