EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Page 200

Endress+Hauser Proline Prosonic Flow P 500
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data Proline Prosonic Flow P 500 HART
200 Endress+Hauser
Fault mode Choose from:
Actual value
0 Hz
Definable value between: 2 to 12 500 Hz
Switch output
Fault mode Choose from:
Current status
Open
Closed
Relay output
Failure mode Choose from:
Current status
Open
Closed
Local display
Plain text display With information on cause and remedial measures
Backlight Red lighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Interface/protocol
Via digital communication:
HART protocol
Via service interface
CDI-RJ45 service interface
WLAN interface
Plain text display With information on cause and remedial measures
Web browser
Plain text display With information on cause and remedial measures
Light emitting diodes (LED)
Status information Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
Supply voltage active
Data transmission active
Device alarm/error has occurred
Diagnostic information via light emitting diodes →  162
Low flow cut off The switch points for low flow cut off are user-selectable.