Diagnosis and troubleshooting Proline Prosonic Flow P 500 HART
176 Endress+Hauser
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
833 Electronics temperature
too low
Increase ambient temperature S Warning
1)
834 Process temperature too
high
Reduce process temperature S Warning
1)
835 Process temperature too
low
Increase process temperature S Warning
1)
836 Process pressure above
limit
Reduce process pressure S Alarm
837 Process pressure below
limit
Increase process pressure S Warning
1)
841 Flow velocity too high Reduce flow rate S Warning
1)
842 Process value below limit Low flow cut off active!
Check low flow cut off configuration
S Warning
1)
870 Measuring inaccuracy
increased
1. Check process
2. Increase flow volume
F Alarm
1)
881 Signal to noise ratio too
low
1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
F Alarm
882 Input signal faulty 1. Check input signal parameterization
2. Check external device
3. Check process conditions
F Alarm
930 Sound velocity too high 1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
S Warning
1)
931 Sound velocity too low 1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
S Warning
1)
941 API/ASTM temperature
out of specificat.
1. Check process temperature with
selected API/ASTM commodity
group
2. Check API/ASTM-related
parameters
S Warning
1)
942 API/ASTM density out of
specification
1. Check process density with selected
API/ASTM commodity group
2. Check API/ASTM-related
parameters
S Warning
1)
943 API pressure out of
specification
1. Check process pressure with selected
API commodity group
2. Check API related parameters
S Warning
1)
953 Asymmetry noise signal
too high path 1 to n
1. Check process conditions
2. Clean or replace transducers
3. Replace sensor electronic module
(ISEM)
M Alarm
954 Sound velocity deviation
too high
1. Check medium configuration
2. Check process conditions
3. Clean or replace transducers
S Warning
1)
1) Diagnostic behavior can be changed.