Proline Prosonic Flow W 400 HART Commissioning
Endress+Hauser 103
Navigation
"Setup" menu → Low flow cut off
‣
Low flow cut off
Assign process variable (1837)
→ 103
On value low flow cutoff (1805)
→ 103
Off value low flow cutoff (1804)
→ 103
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign process variable – Select process variable for low
flow cut off.
• Off
• Volume flow
• Mass flow
• Flow velocity
Flow velocity
On value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→ 103).
Enter on value for low flow cut
off.
Positive floating-
point number
0.3 m/s
Off value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→ 103).
Enter off value for low flow cut
off.
0 to 100.0 % –