Notes on use
10
Entering the menu
The navigation always starts from the main screen (measured value display
2)
). From there, the fol-
lowing menus can be opened by the keys:
L00-FMU90xxx-19-00-00-yy-038
• shortcut menu
The shortcut menu is accessed via the "Info" key. It allows quick access to device information:
– daily counter (for flow measurements)
–tag marking
– envelope curve: used to check the signal quality
– language: sets the display language
– device information: serial number, versions of software and hardware
– password/reset: used to enter the password or reset code
All parameters of the shortcut menu are contained in the main menu as well.
• main menu
The main menu is accessed via the "Menu" key. It contains all parameters of the Prosonic S. It
is divided into submenus. Some of the submenus consist of further submenus. Which submenus
are actually present, depends on the instrument version and the installation environment.
An overview of all submenus and parameters is given in the chapter "Operating menu".
• actual error
If the self-monitoring of the Prosonic S detects an error, the softkey symbol appears
above the middle key.
If the softkey symbol flashes, only "warnings" are present.
If the softkey symbol is displayed permanently, at least one "alarm" is present.
After pressing the key, a list of all currently present errors appears.
2) Note: Depending on the configuration, the appearance of the measured value display may be differemt from the example in the figure.
shortcut menu main menu
actual error
daily counter*
level
Warning 01802
flow
Alarm 01502
tag marking
envelope curve
language
device information
password/reset
* for flow measurements only
...
...
...
...
1
2
3
4
5
6
FMU90