The "system information" menu
150
9.7.4 "application parameter"
"sensor N" (N = 1 or 2)
Indicates if a setting which depends on the application parameters ("tank shape", "medium
property", "process condition") has been changed after the setting of the application parameters in
the service menu.
9.7.5 "echo quality sensor"
"echo quality N" (N = 1 or 2)
Displays the echo quality of the respective sensor.
The echo quality is the distance (in dB) between the level echo and the Floating Average Curve
(FAC).
appl. parameter E1405
sensor 1:
sensor 2:
echo quality sen. E1406
echo quality 1:
echo quality 2: