Loading...
The "level" menu
18
2 The "level" menu
2.1 The "basic setup" submenu
2.1.1 "LVL N sensor selection" (N = 1 or 2)
"input"
Use this parameter to assign a sensor to the channel.
Selection:
•no sensor
•sensor 1
sensor 2 (only for 2-channel instruments)
"sensor selection"
Use this parameter to specify the type of the connected ultrasonic sensor.
!
Note!
For the sensors FDU9x the option "automatic" is recommended (default setting). With this setting
the Prosonic S recognizes the type of sensor automatically.
For the sensors FDU8x the type has to be assigned explicitly. The automatic sensor recognition
does not work for these sensors.
"
Caution!
After exchanging a sensor, observe the following:
The automatic sensor recognition is also active after a sensor has been exchanged
3)
. The Prosonic S
recognizes the type of the new sensor automatically and changes the "detected" parameter if
required. The measurement continues without a break.
Nevertheless, in order to ensure perfect measurement, the following checks are required:
•Check the "empty calibration" and "full calibration" parameters. Adjust these values if
required. Take into account the blocking distance of the new sensor.
•Go to the "distance correction" parameter set and check the displayed distance. If required,
perform a new interference echo suppression.
"detected" (only available for "sensor selection" = "automatic")
Indicates the type of the automatically detected sensor.
level LX001
level (LVL) 1
level (LVL) 2
"level" selection list
Use this list to select the level channel you are going to configure.
LVL1 sensor sel. L1003
input:
sensor selection:
detected:
3) if the new sensor is of the type FDU9x

Table of Contents