EasyManua.ls Logo

Endress+Hauser Prosonic S FMU90 - ISO-Venturi Flumes

Endress+Hauser Prosonic S FMU90
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
185
13.1.2 ISO-Venturi flumes
L00-FMU90xxx-05-00-00-xx-004
BD: blocking distance of the sensor
!
Note!
After selecting the type of flume, Q
max
can be adjusted to the flow conditions. Q
max
defines the flow
at which the output current is 20 mA.
b
0
H
max
b
e
b
0
H
b
e
BD
³ 3 H
max
Type of flume b
0
[mm] b
e
[mm] H
max
[mm] Q
max
[m
3
/h]
ISO-Venturi 415 150 75 200 42,5
ISO-Venturi 425 250 125 300 130,3
ISO-Venturi 430 400 200 400 322,2
ISO-Venturi 440 400 267 625 893,6
ISO-Venturi 450 500 333 700 1318,9
ISO-Venturi 480 800 480 800 2200