The "device properties" menu
136
8 The "device properties" menu
8.1 The "operating parameters" submenu
8.1.1 "distance unit"
Defines the distance unit.
Selection:
•m (default)
•ft
•mm
•inch
8.1.2 "temperature unit"
Defines the temperature unit.
Selection:
•°C (default)
•°F
8.1.3 "operating mode"
Use this parameter to select the operating mode. The available selection depends on the instrument
version.
Selection:
level
level + flow
15)
•flow
15
flow + backwater
15,16)
distance unit D1101
distance unit:
temperature unit D110B
temperature unit:
operating mode D110C
operating mode:
15) only for instruments with flow measurement software (FMU90 - *2********** and FMU90-*4**********)
16) only for instruments with 2 sensor inputs