Appendix
186
13.1.3 British standard Venturi flumes (BS 3680)
L00-FMU90xxx-05-00-00-xx-005
BD: blocking distance of the sensor
The bottom of the flume may not have any slope throughout the length x. (no measuring flume with data threshold)
!
Note!
After selecting the type of flume, Q
max
can be adjusted to the flow conditions. Q
max
defines the flow
at which the output current is 20 mA.
2 H
max
3 H
max
3 (B-b)
b
B
L
BD
x
H
max
Type of flume b H
max
[mm] Q
max
[m
3
/h]
BST Venturi 4" 4" 150 36,25
BST Venturi 7" 7" 190 90,44
BST Venturi 12" 12" 340 371,1
BST Venturi 18" 18" 480 925,7
BST Venturi 30" 30" 840 3603

Table of Contents