The "relays/controls" menu
75
5.1.4 relay N (N = 1 - 6)
(Parametrization of a time pulse relay)
(only for instrument versions with flow functionality:
FMU90 - *2********** and FMU90 - *4**********)
"pulse width" and "pulse time"
5.1.5 "relay N" (N = 1 - 6)
(Parametrization of a counting pulse relay)
"counter unit"
Use this parameter to select the unit for the flow volume.
Selection:
•l (default)
•hl
•Ml
•m
3
•dm
3
•cm
3
•ft
3
•inch
3
•us gal
•us mgal
•i gal
•barrels
relay 2 R2103
function: time pulse
pulse width:
pulse time:
Use these parameters to specify the time interval
between two pulses (pulse time) and the
duration of each pulse (pulse width).
pulse time
•unit: min
•default: 1 min
range of values: 1 to 65000 min
pulse width
•unit: ms
default: 200 ms
range of values: 200 to 60000 ms
L00-FMU90xxx-19-00-00-yy-063
A: pulse time; B: pulse width;
C: relay energized; D: relay de-energized
t
C
D
A
B
relay 2 R2103
function: pulse flow 1
counter unit:
pulse value:
pulse width: