EasyManua.ls Logo

Endress+Hauser Prosonic S FMU90 - Number of Pumps; Standby Pump; Reset

Endress+Hauser Prosonic S FMU90
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The "relays/controls" menu
99
"pump control N" (N = 1 or 2)
"reference"
Defines the level channel to which the pump control refers.
Selection:
none (default)
level 1
level 2 (for instrument versions with 2 level inputs)
"number of pumps"
Defines the number of pumps participating in the pump control. At the end of the configuration
procedure a relay must be allocated to each of the pumps ("relay allocation" parameter set).
Range of values : 1 ... 6 (depending on the number of relays)
•Default: 1
"standby pump"
(only for instruments with external limit switches)
Determines if one of the pumps is a standby pump.
Selection:
no (default)
There is no standby pump.
yes
The last of the pumps is a standby pump. It can not be configured individually. If a failure of one
of the other pumps is reported to the Prosonic S, the standby pump substitutes for this pump.
Example:
number of pumps: 5
standby pump: yes
=> pump control for pumps 1 to 4; pump 5 is the standby pump
"reset"
This parameter is used to restart the pump control (e.g. after a failed pump has been repaired).
!
Note!
The reset has the same effect as a disconnection of the supply voltage. It does not influence the
parametrization of the pump control.
Selection:
no (default)
The pump control is not reset.
yes
The pump control is reset.
pump control R1401
reference:
number of pumps:
standby pump:
reset:

Table of Contents

Other manuals for Endress+Hauser Prosonic S FMU90

Related product manuals