Notes on use
5
Display symbols
LEDs
Symbol Meaning
Operating mode of the instrument
User
User parameters can be edited. Service parameters are locked.
Diagnosis
The service interface is connected.
Service
User and service parameters can be edited.
Locked
All parameters are locked.
Locking state of the currently displayed parameter
Display parameter
The parameter can not be edited in the current operating mode of the instrument.
Editable parameter
The parameter can be edited.
Scroll symbols
Scroll list available
Indicates that the list contains more parameters than can be represented on the display. By pressing
V or W repeatedly, all parameters of the list can be accessed.
Navigation in the envelope curve display
Move left
Move right
Zoom in
Zoom out
LED indicating the operating state (pos. (e) in the figure)
green normal measuring mode; no error detected
red (flashing)
Warning:
An error is detected but the measurement continues. Reliability of the measured value is no longer
ensured.
red
Alarm:
An error is detected. The measurement is interrupted. The measured value assumes the value
specified by the user (parameter "output on alarm").
off supply voltage missing
LEDs for the relays (pos. (f) in the figure)
yellow The relay is activated.
off The relay is de-activated (idle state).