Section 9: Verifying the Model 9500B Accuracy Specification 9-5
Final Width = 215mm
Verif. Output Absolute Tolerance Measured
Point. Voltage Limits (-DCV) Value
Lower Higher
-DC1 1.0000mV 0.97475mV 1.02525mV
-DC2 1.9000mV 1.87453mV 1.92547mV
-DC3 2.3000mV 2.27443mV 2.32557mV
-DC4 5.0000mV 4.97375mV 5.02625mV
-DC5 6.0000mV 5.9735mV 6.0265mV
-DC6 19.000mV 18.9703mV 19.0297mV
-DC7 23.000mV 22.9693mV 23.0307mV
-DC8 50.000mV 49.9625mV 50.0375mV
-DC9 60.000mV 59.960mV 60.040mV
-DC10 190.00mV 189.928mV 190.073mV
-DC11 230.00mV 229.918mV 230.082mV
-DC12 500.00mV 499.85mV 500.15mV
-DC13 600.00mV 599.83mV 600.18mV
-DC14 1.9000V 1.8995V 1.9005V
-DC15 2.3000V 2.2994V 2.3006V
-DC16 5.0000V 4.99873V 5.00127V
-DC17 6.0000V 5.99848V 6.00152V
-DC18 19.000V 18.99523V 19.00477V
-DC19 23.000V 22.99423V 23.00577V
-DC20 50.000V 49.98748V 50.01252V
-DC21 60.000V 59.98498V 60.01502V
-DC22 190.00V 189.9525V 190.0475V
Table 9.8.1.2 — DC/Square DC (-) Verification into 1M