7-10 Section 7: Model 9500B — Specifications
Final Width = 215mm
7.10 Trigger Output Specifications
Amplitude: 1Vpk-pk into 50
Rise and Fall Time: < 1ns
Aberrations: < ±10%
Source VSWR: 1.2:1 typ
7.10.1 Input Leakage (Not available via 9550 Active Head)
Function Signal Trigger Waveform Trigger Frequency Trigger Alignment Trigger Leads by
Voltage DC Square 100Hz
Square Square Signal div 1/10/100 Depart from ground 1/64 Period
Levelled Sine <100MHz Square Signal Rising Zero Cross
>100MHz Square Signal div 1/10/100 Rising Zero Cross
>3.2GHz Square Signal div 2/20/200 Rising Zero Cross
Edge 500ps 1:9 Pulse Signal div 1/10/100 Return to Ground 25ns
HV Square Signal div 1/10/100 Depart from ground 1/64 Period
150ps 1:9 Pulse Signal div 1/10/100 Return to Ground 25ns
25ps 1:9 Pulse Signal div 1/10/100 Return to Ground 25ns
70ps 1:9 Pulse Signal div 1/10/100 Return to Ground 25ns
Timing Markers (>10ns) Normal 1:19 Pulse Signal div 1/10/100 Rising Edge or Triangle Peak
Highlighted 1:199 Pulse Signal div 1/10/100 Rising Edge or Triangle Peak
Current DC Square 100Hz
Square Square Signal div 1/10/100 Depart from ground 1/64 Period
Composite Video Line Composite Sync Line Freq Sync lead edge
Field Pulse Frame Freq Frame start
Linear Ramp 1:2 Pulse Signal div 1/10/100 Start of rise or fall
Overload Pulse Step or Square Single or 100Hz Lead Edge
Input Leakage, Auxiliary Input and Input Impedance functions have no related trigger output. Free running 100Hz signal provided
No trigger provided in Pulse Width function