DTX Series CableAnalyzer
Technical Reference Handbook
3-28
1
2
3
4
5
6
7
8
amd13f.eps
A The location of the ELFEXT results. Press J to switch
between the tester and smart remote.
B The overall ELFEXT result. “PASS*/FAIL* Results” on page
3-12 describes results marked with an asterisk.
C Magnification level for the plot. Use AD to zoom
in or out at the cursor’s location.
D Measured ELFEXT for the cable pairs. Higher ELFEXT means
better cabling performance.
E The limit line (in red) for ELFEXT. The higher the
measurements rise above the limit line, the better the
cabling performance. Press
L to see plots of individual
pairs.
F The cursor and its location on the frequency scale. When
you first view the plot, the cursor is placed at the worst
margin. Use
BC to move the cursor.
G The measured ELFEXT and margin at the cursor’s position.
Margin is the difference between the measured value and
the limit. Margin is negative if the pair failed.
H The horizontal scale is the frequency range in megahertz.
The vertical scale is the ELFEXT range in decibels.
Figure 3-21. ELFEXT Plot