Certifying Twisted Pair Cabling
Twisted Pair Test Settings 3
3-3
Twisted Pair Test Settings
Table 3-1 describes the settings that apply to twisted pair
cabling tests.
To access the settings, turn the rotary switch to SETUP,
use
D to highlight Twisted Pair; then press H.
Table 3-1. Twisted Pair Test Settings
Setting Description
SETUP >
Twisted Pair >
Cable Type
Select a cable type appropriate for the type you will test. The cable types are organized by type and
manufacturer.
SETUP >
Twisted Pair >
Test Limit
Select the appropriate test limit for the job.
SETUP >
Twisted Pair >
NVP
Nominal velocity of propagation, which is used with the measured propagation delay to determine
cable length. The default value defined by the selected cable type represents the typical NVP for that
cable type. You may enter a different value if necessary. To determine the actual value, change the NVP
until the measured length matches the known length of a cable. Use a cable at least 15 m (50 ft) long.
The recommended length is 30 m (100 ft).
Increasing the NVP increases measured length.
-continued-