DTX Series CableAnalyzer
Technical Reference Handbook
9-14
Specifications
Specifications apply at 23
o
C (73
o
F), unless otherwise
noted.
Environmental and Regulatory Specifications
Operating temperature 32 °F to 104 °F (0
o
C to 40
o
C)
Storage temperature -4 °F to +140 °F (-20
o
C to +60
o
C)
Operating relative humidity
(% RH without condensation)
32 °F to 95 °F (0 °C to 35 °C): 0 % to 90 %
95 °F to 113 °F (35 °C to 45 °C): 0 % to 70 %
Vibration Random, 2 g, 5 Hz-500 Hz
Shock 1 m drop test with and without module and adapter attached
Safety CSA C22.2 No. 1010.1: 1992
EN 61010-1 1
st
Edition + Amendments 1, 2
Pollution degree 2
Altitude Operating: 4000 m; Storage: 12000 m
EMC EN 61326-1
Laser safety (for fiber test module) Class I CDRH. Complies to EN 60825-2