Loading...
DTX Series CableAnalyzer
Technical Reference Handbook
5-44
Using FindFiber in Loopback Mode
Use FindFiber in Loopback mode to quickly check the
continuity of patch cords and fiber spools. Figure 5-22
shows the equipment needed for using FindFiber in
Loopback mode.
1 Turn the rotary switch to SETUP; then select Fiber.
2 Select Remote End Setup; then select Loopback.
3 Clean all connectors; then connect the tester's
OUTPUT fiber to one end of the fiber path, as shown
in Figure 5-23.
4 Turn the rotary switch to MONITOR; then select
FindFiber.
5 Try various connections with the INPUT fiber. The
status shows Loopback for both fibers when the
path is complete.
Wi
p
e
1
2
3
4
5
TALK
MONITOR
ENTER
TEST
SAVE
SPECIAL
FUNCTIONS
SETUP
AUTO
TEST
SINGLE
TEST
EXIT
F1 F2 F3
amd55f.eps
A Tester with fiber module
B AC adapters with line cord (optional)
C Two patch cords (match fiber and connectors to be
tested, SC at tester end)
D Two adapters of the appropriate type
E Fiber cleaning supplies
Figure 5-22. Equipment for Using FindFiber in Loopback Mode

Table of Contents