EasyManuals Logo

GSi TYMPSTAR Series User Manual

GSi TYMPSTAR Series
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #133 background imageLoading...
Page #133 background image
Calibration
Step 8
Adjust the overall calibration output level if necessary by pressing the Cal HL up/
down softkeys. The level can be adjusted up or down in 5.0dB increments.
Step 9 (Custom Mode only)
When a custom transducer is being calibrated as discussed earlier in Custom Trans-
ducers, the TARG SPL softkeys will be active.
Adjust the target SPL requirement for each frequency by pressing the TARG
SPL up/down softkeys.
The level can be adjusted up or down in 0.5dB incre-
ments
.
The final values must be within 10dB of the GSI defaults.
Step 10
Follow the instructions on the screen until all stimulus calibrations are complete.
Step 11
Press the RETURN hardkey twice to return to the Main Calibration Menu.
GSI TympStar Version 1 and Version 2 Service Manual 4 - 61

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GSi TYMPSTAR Series and is the answer not in the manual?

GSi TYMPSTAR Series Specifications

General IconGeneral
BrandGSi
ModelTYMPSTAR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals