Operation Summary
Changing Instrument Versions
Changing from a The Version function provides an easy way for a TympStar Version 2 to be
Version 2 to a changed to a Version 1 instrument either temporarily or permanently. Version
Version 1 1 functionality is a limited subset of Version 2 functionality and does not in-
clude 678 and 1000 Hz probe tones, multi-frequency tympanometry or ARLT
testing.
To change from a TympStar Version 2 to a
Version 1:
1) Press the ETF hardkey to select the
ETF test mode, then press the RE-
TURN hardkey to display the ETF sub-
menu.
INTACT PERF
INSTRUMENT
PROGRAM
TM TM OPTIONS MODE
2) Press the INSTRUMENT OPTIONS softkey, then the MORE softkey to dis-
play the VERSION softkey selection.
GSI TympStar Version 1 and Version 2 Service Manual
3 - 11