Calibration
Resolving problems
It is unlikely that you will encounter problems during calibration. However, if you
are unable to meet a calibration objective at any point in the process you must take
corrective action and start the process again from the beginning.
If you encounter a problem-
Check your equipment, test cavities and connections, and start over.
Still have a problem-
Clean the probe or Contra phone, check the probe tubes and start over.
Still have a problem-
Change the probe assembly and start over.
Still have a problem-
Contact your local GSI representative or GSI technical support.
GSI TympStar Version 1 and Version 2 Service Manual 4 - 27