Chapter 4
Step 1
Press the PROBE SPL CAL softkey.
The PROBE TONE SPL CAL screen will be displayed.
Step 2
Follow the instructions on the screen to calibrate probe tone frequencies from 226
Hz through MULT HZ (1500).
Step 3
Press the VERIFY MULT HZ softkey. The VERIFY MULT HZ screen will be
displayed.
Grason-Stadler
Probe Tone SPL Cal
4 - 62