Operation Summary
1) Select the ETF hardkey followed by the RETURN hardkey.
2) Select the INSTRUMENT OPTIONS softkey and then press the TEST
SEQUENCE softkey.
3)
Select the REFLEX THRESHOLD softkey and press the ON/OFF softkey
to activate or deactivate reflex threshold.
4)
Acoustic reflex measurements can be obtained ipsilaterally or
contralaterally.
Select the STIMULUS EAR softkey to choose from
Ipsi only, Contra only, or Ipsi and Contra for reflex testing.
5) Press the STIMULUS Hz softkey to select the stimulus to be included in
the Reflex Threshold Auto Sequence
.
Scroll through each stimulus by
pressing the up and down arrow softkeys to scroll through the stimuli and
toggle the YES/NO softkey to program each stimulus for Reflex
Threshold testing in the auto test sequence.
6) Select the REFLEX DECAY softkey and press the ON/OFF softkey to
activate or deactivate reflex decay.
7) Select the STIMULUS Hz softkey to select the stimuli for Reflex Decay
testing. Press the up and down arrow softkeys to scroll through the
stimuli and toggle the YES/NO softkey to select or de-select the
stimulus.
8) Press the RETURN hardkey twice and press the STORE softkey to
store test selections in memory.