Chapter 8
INVALID OUTPUT VOLTAGE LIMIT REACHED
An attenuator level was selected in the calibration mode using the set SPL up/down
or intensity up/down keys for a stimulus that is outside of the valid attenuator limits.
INVALID SELECTION CUSTOM REFERENCE THRESHOLD MODE
NOT SELECTED
An attempt was made to modify the SPL limits using the set SPL up/down softkeys
with the custom reference threshold mode not selected.
WARNING INSUFFICIENT WARM UPTIME—UNIT MUST BE ON 10
MINUTES
The unit has not sufficiently warmed up (10 minutes) in the calibration mode. The
calibration may be performed, but the accuracy is not guaranteed.
INVALID SPL LIMIT REACHED
An attempt was made to modify the set SPL value to more than ±10 dB HL away
from the GSI set SPL value for that stimulus.
NR <>
Screening reflex test result indicating that the test was stopped for that stimulus with
no reflex detected due to an attenuator limit error when setting the attenuator for the
intensity being tested.
NT <>
Screening reflex test result indicating that no test was performed for that stimulus
due to an attenuator limit error when setting the attenuator for the intensity being
tested.
NOTE GSI DEFAULTS USED - REPROGRAM
Try reprogramming the user key data.
NOTE TEST NOTAVAILABLE - CONTACT SERVICE
Calibration should be performed on the IPSI and CONTRA transducers.
NOTE REFLEX DECAY NOTAVAILABLE - CONTACT SERVICE
Calibration should be performed on the IPSI and CONTRA transducers.
ERROR INVALID SELECTION DURING DIAGNOSTIC MODE
The key pressed is not valid while in the diagnostic mode.
PROBE TONE LEVEL CHANGED - PERFORM Y CALIBRATION
The probe tone SPL calibration has been changed requiring a recalibration of the
compliance.