REMOTE CONTROL
131
OT (Over Temperature)
Over temperature protection has
been tripped
4
16
OCP
OCP protection has been tripped
5
32
Output Short
Output Short protection has been
tripped
6
64
System Wait
If output short protection has
been tripped, it requires to wait
10s for output on again after
clearing protection
7
128
CAL (Calibration data is invalid)
8
256
OP (Over-Power)
Over power protection has been
tripped
9
512
Always 0
15
32768
Condition
Register
The Questionable Status Condition Register
indicates the status of the power supply. If a bit is
set in the Condition register, it indicates that the
event is true. Reading the condition register does
not change the state of the condition register.
PTR/NTR Filters
The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will set the corresponding bit in the
Event Registers. Use the Positive transition filter to
view events that change from false to positive, and
use the negative transition filter to view events
that change from positive to negative.
Positive Transition
01
Negative Transition
10