REMOTE CONTROL
63
:MEASure[:SCALar]:POWer[:AC]:APParent
Query
Description
Returns the apparent power (VA).
Syntax
:MEASure[:SCALar]:POWer[:AC]:APParent?
Return
<NR2>
Returns the apparent power in VA.
:MEASure[:SCALar]:POWer[:AC]:PFACtor
Query
Description
Returns the power factor.
Syntax
:MEASure[:SCALar]:POWer[:AC]:PFACtor?
Return
<NR2>
Returns the power factor.
:MEASure[:SCALar]:POWer[:AC]:REACtive
Query
Description
Returns the reactive power (VAR).
Syntax
:MEASure[:SCALar]:POWer[:AC]:REACtive?
Return
<NR2>
Returns the reactive power in VAR.
:MEASure[:SCALar]:POWer[:AC][:REAL]
Query
Description
Returns the active power in Watts.
Syntax
:MEASure[:SCALar]:POWer[:AC][:REAL]?
Return
<NR2>
Returns the power in W.
:MEASure[:SCALar]:VOLTage[:RMS]
Query
Description
Returns the voltage (Vrms).
Syntax
:MEASure[:SCALar]:VOLTage[:RMS]?
Return
<NR2>
Returns the voltage in Vrms.