APS-7000 Programming Manual
90
:SYSTem:WRELease
Set
Description
Clears the protection circuits (OCP, OTP). This
command is functionally the same as
the :OUTPut:PROTection:CLEar command (page
66).
Syntax
:SYSTem:WRELease
:SYSTem:IPKHold:TIME
Set
Query
Description
Sets or queries the Ipeak hold time for peak current
measurement when output on.
Syntax
Query Syntax
:SYSTem:IPKhold:TIME {<NR1>}
:SYSTem:IPKhold:TIME?
Parameter
<NR1>
1~60,000
Example
:SYSTem:IPKHold:TIME 10
Sets the Ipeak hold time 10ms to measure when
output on.
:SYSTem:HOLD:STATe
Set
Query
Description
Sets or queries the freeze hold state on/off.
Syntax
Query Syntax
:SYSTem:HOLD:STATe {<bool>|OFF|ON}
:SYSTem:HOLD:STATe?
Parameter
OFF | 0
Turns the freeze hold off.
ON | 1
Turns the freeze hold on.
Return parameter
<bool>
Returns the freeze hold status.