REMOTE CONTROL
71
:STATus:QUEStionable:PTRansition
Set
Query
Description
Sets or queries the bit sum of the positive
transition filter of the Questionable Status register.
Syntax
Return Syntax
:STATus:QUEStionable:PTRansition <NR1>
:STATus:QUEStionable:PTRansition?
Parameter
<NR1>
0~32767
Return parameter
<NR1>
0~32767
:STATus:PRESet
Set
Description
This command resets the ENABle register, the
PTRansistion filter and NTRansistion filter on the
Operation Status, Questionable Status and
Warning Status Registers. The registers/filters will
be reset to a default value.
Default Register/Filter Values
Setting
QUEStionable Status Enable
0x0000
QUEStionable Status Positive Transition
0x7FFF
QUEStionable Status Negative Transition
0x0000
Operation Status Enable
0x0000
Operation Status Positive Transition
0x7FFF
Operation Status Negative Transition
0x0000
WARNing Status Enable
0x0000
WARNing Status Positive Transition
0x7FFF
WARNing Status Negative Transition
0x0000